The content is as follows:

12-1JULY 2004

MARLAP

Sampling

Concentration, Separation,

Isolation, etc. Steps

Sample

Preparation

Measurement

FIGURE

12.1ŠDegree of error in laboratory sampl

e preparation relative to other activities

12 LABORATORY SAMPLE PREPARATION12.1Introduction

On first impression, sample preparation may s

eem the most routine aspect of an analytical

protocol. However, it is critical that analysts

realize and remember that

a measurement is only as

good as the sample preparation

that has preceded it. If an a

liquant taken for analysis does not

represent the original sample accurate

ly, the results of this analysis

are questionable. As a general

rule, the error in sampling and the sample

preparation portion of an analytical procedure is

considerably higher than that in the me

thodology itself, as illu

strated in Figure 12.1.

One goal of laboratory sample

preparation is to provide, without sample loss, representative

aliquants that are free of laboratory contaminati

on that will be used in the next steps of the

protocol. Samples are prepared in accordance

with applicable standard operating procedures

(SOPs) and laboratory SO

Ps using information provided by fi

eld sample prepar

ation (Chapter 10,Field and Sampling Issues that Affect Laboratory Measurements

), sample screening activities,

and objectives given in the appropriate pl

anning documents. The laboratory sample preparation

techniques presented in this chapter include the

physical manipulation of the sample (heating,

screening, grinding, mixing, etc.) up to the

point of dissolution. Steps such as adding

carriers and tracers, followed by wet ashing or

fusion, are discussed in Chapter 13 (SampleDissolution) and Chapter 14 (SeparationTechniques).This chapter presents some general guidance

Contents

12.1Introduction

……………………..12-1

12.2General Guidance for Sample Preparation..12-2

12.3Solid Samples

…………………..12-12

12.4Filters

………………………..12-30

12.5Wipe Samples

…………………..12-31

12.6Liquid Samples

………………….12-32

12.7Gases

…………………………12-36

12.8Bioassay

………………………12-36

12.9References

……………………..12-37